02-557-1073

The LogTag® Dry Ice “Probe-less” Temperature Logger operates, measures and stores up to 8000 temperature readings in temperature environments ranging from -80°C to +40°C (-112°F to +104°F).

 

Manufacturer:
LogTag
Availability:
In Stock
SKU:
TL-007-Logtag

The LogTag® Dry Ice “Probe-less” Temperature Logger operates, measures and stores up to 8000 temperature readings in temperature environments ranging from -80°C to +40°C (-112°F to +104°F).

 

Intended for use in transit monitoring of articles stored in packaging incorporating dry ice cooling agents.

 

SPECIFICATIONS

Product Model TRIL-8.
Sensor Measurement Range -80°C to +40°C (-112°F to +104°F).
Operating Temperature Range -80°C to +40°C (-112°F to +104°F).
Storage Temperature Range -20°C to +40°C (-4°F to +104°F).
Rated Temperature Reading Accuracy Better than ±1°C (±1.8°F) for -30°C to +20°C (-22°F to +68°F).
Better than ±1.2°C (±2.1°F) for -45°C to -30°C (-49°F to -22°F) and +20°C to +40°C (+68°F to +104°F).
Better than ±1.7°C (±3.1°F) for -80°C to -45°C (-112°F to -49°F).
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration.
Rated Temperature Reading Resolution Less than 0.1°C (0.2°F) for -80°C to 0°C (-112°F to +32°F).
Less than 0.2°C (0.4°F) for 0°C to +20°C (+32°F to +68°F).
Less than 0.5°C (0.9°F) for +20°C to +40°C (+68°F to +104°F).
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
Sensor Reaction Time Typically less than 5 minutes (T90) in moving air (1m/s).
Recording Capacity 8031 temperature readings. 53 days @ 10min logging, 80 days @ 15min logging.
Sampling Interval Configurable from 1 minute to several hours.
Logging Start Options Push button start or specific date & time.
Recording Indication Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator.
Download Time Typically less than 5 seconds for full memory (8031 readings), depending on computer or readout device used.
Environmental IP65 (roughly equivalent to NEMA 4).
Power Source 3.6V low temperature chemistry Lithium (Fixed) battery.
Battery Life Minimum storage life of 12 months before ‘start’.
Rated for up to a total accumulative exposure of 1000hrs @ -80°C across multiple trips.
Real Time Clock Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Size 86mm(H) x 54.5mm(W) x 8.6mm(T).
Weight 33g.
Case Material Polycarbonate.